Vue de la page 0
HB3121
1
USER MANUAL
COPYRIGHT
The copyright in this document which contains information is vested in
CASELLA LIMITED. The contents of this document must not be used for
purposes other than that for which it has been supplied or reproduced or disclosed
wholly or in part without the prior written permission of CASELLA LIMITED.
CASELLA LIMITED
Regent House Telephone: +44 (0)1234 841441
Wolseley Road Facsimile: +44 (0)1234 841490
Kempston Telex: 827707
Bedford e-mail: [email protected]
MK42 7JY Webpage: http://www.casella.co.uk
FORTIN
BAROMETER
HB3121-02
Vue de la page 0
1 2 3 4 5 6 ... 10 11

Résumé du contenu

Page 1 - BAROMETER

HB31211USER MANUALCOPYRIGHTThe copyright in this document which contains information is vested inCASELLA LIMITED. The contents of this

Page 2

HB312110MAINTENANCENo attention should be necessary to a barometer except to dust it and to keep theadjusting screw clean and lightly oiled. The thre

Page 3

HB312111SPECIFICATIONEffective range: Ordinary Range model900 to 1060mb670 to 800mmHg26.4 to 31.2inHgAltitude up to: 450mVernier reads to: 0.1mb, 0.05

Page 4

HB31212CONTENTS PageINTRODUCTION... 3PRINCIPL

Page 5

HB31213INTRODUCTIONThe Fortin barometer is the most widely used type and can be found in manyindustrial laboratories, workshops, classrooms and meteor

Page 6 - ADJUSTING

HB31214INITIAL SET UPCHOOSING A SUITABLE LOCATIONThe room in which the barometer is due to hang should: -• Not be unduly warm.• Not be subject to va

Page 7

HB31215MOUNTING THE INSTRUMENTHaving chosen the best possible position for the barometer the glass windowedcase should be securely fixed to

Page 8

HB31216If the readings are to be corrected to mean sea level, the height of the cisternshould be accurately found and recorded as

Page 9

HB31217READING THE BAROMETERThe thermometer should always be read first since the presence of the observerwill cause the thermometer to al

Page 10

HB31218Where to view the vernier from

Page 11

HB31219The height of the mercury column can now be read on the scale using the vernier.The diagram below gives an example.For mmHg the reading is 1000