AFC123 Personal Sampler. Page 22
Handbook No. HB3182-02 Issue2/JULY 1994
Symptom:- Pump runs but cannot achieve desired flow rate.
Possible cause:-
Service Hint.
Leak on pump assembly. Locate and correct.
Dirt under valves or valves damaged. Replace.
Inlet filter clogged. Remove and clean.
Flow potentiometer inoperative Replace circuit
board.
Symptom:- Cannot obtain full shift
working even when pressure drop does
not exceed 40cm H20 at 2 litres/min.
Possible cause:-Incomplete charge. Recharge.
Battery below capacity. Fully charge and
check voltage, should
be 4.8V-5.4V. If
below check output from charger:-
should be 110-125mA.
If output correct
replace battery.
High current consumption due to Locate and correct.
mechanical defect in pump.